Recall Z-2307-2008 ยท JAK System, X-Ray, Tomography, Computed

Philips And Neusoft Medical Systems Co., Ltd.

NeuViz Dual series computed Tomography Scanner System, X-Ray System

DateSep 17, 2008 (initiated Jun 23, 2008)
StatusTerminated
ReasonPotential for R-host box and components inside to be detached of its mounting during the gantry rotor rotation. There is a potential for components to be damaged inside the system gantry and for the internal components to be expelled from the system gantry and injure the patient, operator, bystander or service people.
Root causeNonconforming Material/Component
ActionOn 6/23/08/08, the Philips Neusoft sent a letter dated 6/10/08 to their US customers explaining the problem, under what circumstances it can occur, and the actions that should be taken by the customer/user in order to prevent risks for patients or users. The field correction is install a R-host bracket to aid supporting of the R-host computer. The letter included additional information to customers who had already received the field action (this was an on-going activity that was re-classified as a potential safety). Customers did not receive the additional information about the status of
Quantity15 units in US
DistributionThe units were distributed to one distributor in NC, and further distributed to medical facilities located in CO, FL, KY, NC, SC, TN and TX.
FDA recordOpen on fda.gov