Recall Z-0071-2008 ยท BTA Pump, Portable, Aspiration (Manual Or Powered)

Neomedix Corp

NeoMedix Trabectome I/A Console, Model Number 550014, surgical aspirator

DateSep 29, 2008 (initiated May 28, 2007)
StatusTerminated
ReasonFailure of irrigation flow: The pinch valve in a few pumps may not consistently open to allow irrigation flow. In over 450 surgeries, the firm confirmed one instance of a valve that did not open. The electrically activated pinch valve in the NeoMedix Trabectome I/A pump activates irrigation flow to the eye during Trabectome surgery.
Root causeNonconforming Material/Component
ActionOn May 31, 2007 NeoMedix mailed a Customer letter to all consignees informing them of the hazard. The letter also indicated to the customers they NeoMedix has developed a minor modification/upgrade that is designed to assure that the pinch valve opens at the appropriate time. The customers were instructed this modification requires return of their existing pump for modification and testing at the firm's facility before its return. In the event that surgeries have already been scheduled, NeoMedix will provide a loaner unit until the suspect unit can be modified and returned to you. The lette
Quantity7
DistributionNationwide to: CA, HI, IA, NY & PA
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K040239SUPRAFLOW CONSOLE
Neomedix Corp., Feb 17, 2004