Recall Z-2301-2008 ยท HWA Impactor

Zimmer Inc.

Zimmer NEXGEN Complete Knee Solution MIS Total Knee Procedure Tibial Broach Impactor (a pounding instrument used during knee surgery), nonsterile, Zimmer, Warsaw, Indiana; REF 5951-90..

DateSep 20, 2008 (initiated Jun 30, 2008)
StatusTerminated
ReasonThe instrument may fracture during use, resulting in metal fragments being left in the patient post-surgery, which could cause implant failure.
Root causeDevice Design
ActionConsignees were notified via letter dated 6/30/08. The letter described the issue, provided instructions for the return of the product and asked customers to complete a reply back form. For assistance, contact Zimmer at 1-574-372-4487.
Quantity1439
DistributionNationwide, Australia, China, England, France, Germany, India, Italy, Japan, Korea, Singapore and Taiwan.