Recall Z-2239-2008 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Abbott Laboratories

ARCHITECT i1000SR System Assay CD ROM WW (excluding US) Special edition Version 1.01 and Version 2.0; list 01P61-01 and 01P61-02. The device is intended to duplicate manual analytical procedures by performing automatically various steps such as pipetting, preparing filtrates, heating, and measuring color intensity.

DateSep 19, 2008 (initiated Jun 16, 2008)
StatusTerminated
ReasonWhen running the Automated Dilution Protocol with the assay files "CMV IgG" or "Toxo IgG" on the ARCHITECT i1000SR system, a software error [9010 (Pipettor Robotics) preventing (Process Path) from operating] is generated and the instrument stops running. No results are generated.
Root causeSoftware design
ActionAbbott affiliates were e-mailed copies of the recall letter Urgent Field Safety Notice Product Correction on 6/16/08 for translation where needed and dissemination to their ARCHITECT accounts who have the affected assay CD-ROMs. Each Abbott international organization will be responsible for contacting customers and determining the method of effectiveness checks based on their country requirements. Abbott sent Product Correction, Immediate Action Required letters dated June 16, 2008 to all active non-U.S. ARCHITECT customers who have the affected assay CD-ROMs. The end users were informed when
Quantity75 CD-ROMs
DistributionAustria, Belgium, Czech Republic, Denmark, France, Germany, Iceland, India, Italy, Latvia, Lithuania, Netherlands, Poland, Portugal, Sweden, Saudi Arabia, Spain, United Arab Emirates