ClearanceBrief
Recall Z-2218-2008 ยท LXH Orthopedic Manual Surgical Instrument

Wright Medical Technology Inc

Dynasty Trial Shell 68mm OD, Group H, REF 3300-GH68, Rx only, Non-Sterile, 1 each, Wright Medical Technology, Inc., 5677 Airline, Road, Arlington, TN 38002 Orthopaedic instrument used for the preparation of the implant site prior to device implantation.

DateSep 17, 2008 (initiated May 20, 2008)
StatusTerminated
ReasonTrials are 2.5mm larger than marked.
Root causeComponent change control
ActionThe firm initiated their recall on May 20, 2008 by issuing a Field Notice to their distributors and followed with telephone calls on May 21 and 22, 2008 requesting the return of the products. The firm then issued letters to their distributors on June 4, 2008 explaining the problem and requesting return of the product from their inventory and from their consignees. The letter included a "fax back" response page to ensure the receipt of the notice. The hospitals were notified of the recall also on June 4, 2008 by means of letters, delivered via Fed Ex, addressed to Hospital Administrators, Ri
Quantity127 units
DistributionWorldwide Distribution --- including USA and Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.