Recall Z-2212-2008 ยท IYN System, Imaging, Pulsed Doppler, Ultrasonic

Siemens Medical Solutions USA, Inc.

Acuson CV70 Ultrasound systems Ultrasonic Pulsed echo/doppler imaging system with Software revisions less than 4.0.0b, material number 784539, 7848521, and 7848547. Is a general purpose, mobile, software-controlled, diagnostic ultrasound system with an on-screen display for thermal and mechanical indices related to potential bioeffect mechanisms.

DateSep 19, 2008 (initiated Jun 4, 2008)
StatusTerminated
ReasonSystem may either display incorrect mechanical and thermal index values, or fail to display them altogether.
Root causeSoftware design
ActionRecall initiated on 06/04/2008. Under the Siemens update program, all consignees will receive a visit from Siemens service personal to install the revised software version. Siemens staff will coordinate with foreign distributors to arrange for service abroad.
Quantity1230 units
DistributionWorldwide Distribution.
FDA recordOpen on fda.gov