Recall Z-2204-2008 ยท NAY System, Surgical, Computer Controlled Instrument

Intuitive Surgical, Inc.

da Vinci S Surgical System, Model number IS2000 Endoscopic Instrument Control System, manufactured by Intuitive Surgical, Sunnyvale, CA.

DateSep 16, 2008 (initiated Jun 1, 2008)
StatusTerminated
ReasonDefective software chip may cause the system to fail and lock up.
Root causeSoftware design (manufacturing process)
ActionRecall initiated June 1, 2008. Urgent Device Recall Letters were sent via e-mail, with follow up phone calls and service visits to correct the product. The firm recommended that continued use of the da Vinci S Surgical System during the time period prior to our service update to replace the software chip. The letter emphasizes the importance of always having backup equipment and instrumentation available when performing surgery with the da Vinci S Surgical System and being prepared to convert to alternative surgical techniques, if necessary. Customers are asked to fill out and return the a
Quantity112 systems
DistributionWorldwide Distribution --- including USA and countries of Belgium, Bulgaria, Canada, China, Denmark, France, Germany, Italy, Korea, Romania, Russia, Saudi Arabia, Singapore, Spain, Sweden, and United Kingdom.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K050369INTUITIVE SURGICAL DA VINCI SURGICAL SYSTEM, MODEL IS2000
Intuitive Surgical, Inc., Apr 29, 2005