Recall Z-2200-2008 ยท HFN Transducer, Pressure, Intrauterine

Utah Medical Products, Inc

Intran Plus, Disposable Intrauterine Pressure Catheter System, Part numbers IUP-400, IUP-400S, IUP-450, IUP-450S, IUP-500, IUP-550, IUP-600, IUP-600S, IUP-650, IUP-650S, IUP-700, IUP-750, Sterile, Single Use, Utah Medical Products Inc., 7043 South 300 West, Midvale, Utah 84047.

DateSep 11, 2008 (initiated Jun 16, 2008)
StatusTerminated
ReasonPossibility of compromised package sterility prior to the expiration date.
Root causeComponent change control
ActionAll consignees were notified by a faxed Recall Notice letter beginning on 6/16/2008. Customers were asked to return all unused product for immediate replacement with units that have a protective barrier within the tray packaging. The firm cautioned customers to visually inspect packaging before use. Follow-up phone calls will be made the following week.
Quantity300,000 units.
DistributionWorldwide Distribution --- USA including states of CA, KS, MO, NC, NV, TX, VA, and WA and country of Japan.