ClearanceBrief
Recall Z-2333-2008 ยท DTB Permanent Pacemaker Electrode

Oscor, Inc.

Permanent Pacing Lead (Catalog #: 03332, 03334, 03337, 03107, 03116, 03108, 03109, 05756, 53420, 05765, 03634, 03633, and 05747); Model #: PY244RU, PY252RU, PY258RU, REFINO58RU, REFINO 52RJU, REFINO 44RJU, REFINO48RJU, BIS/IS-15, BIS/BIS-17, VKU-10V, MP52PV, MP40PV, C/IS-10; Manufactured by Oscor. Pacing and sensing of the ventricle and/or atrium of the heart in conjunction with a compatible, im

DateSep 24, 2008 (initiated Jan 29, 2008)
StatusTerminated
ReasonThe O-rings are over tolerance making it hard to connect the leads to the pacemaker.
Root causeNonconforming Material/Component
ActionOn January 24, 08, Oscor sent out a letter to doctors/ customers with the subject line of "Re: Market Withdrawal of Permanent Pacing Leads due to the lead connector insertion difficulty," stating they experienced some difficulties with the insertion of pacing lead connectors into the pacemaker headers. They asked that customers review the inventory list and remove all products, whose serial numbers match the serial numbers on the list and have not been implanted. Also, customers should call Oscor customer service for a Return Goods Product number and return the product back to the firm for r
DistributionNationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K010787PACING LEAD BIPOLAR ADAPTERS AND EXTENSIONS
Oscor, Inc., May 22, 2001
K040569PERMANENT PACING LEAD, MODEL REFINO, MODEL: REFINO-R, RU, RJU
Oscor, Inc., Apr 7, 2004
K041809PERMANENT PACING LEAD, MODEL PY2
Oscor, Inc., Aug 6, 2004
K872971PERMANENT MYOCARDIAL LEADS MP SERIES
Oscor Medical Corp., Feb 12, 1988