ClearanceBrief
Recall Z-2199-2008 ยท EJB Handle, Instrument, Dental

Biomet 3i, Inc.

Navigator System Surgical kit (SGKIT), handles inside of SGKIT Surgical Kit SGDPH1 through SGDPH5. The product is used for CT guided surgery.

DateSep 16, 2008 (initiated May 9, 2008)
StatusTerminated
ReasonThe drill malfunctions while in use. The drill has been observed to become lodged in the handle.
Root causeNonconforming Material/Component
ActionThe recall was initiated in 5/9/2008. Each US customer was contacted by telephone using a prepared phone script. Following telephone contact, Biomet 3i performed a follow-up of each call with a fax letter. International customers were first contacted by E-mail then by fax letter. Each customer was instructed to check their respective inventory for the affected product and corresponding lot number. They were then requested to fax back a response form to Biomet 3i Regulatory Services to provide feedback concerning their inventory. They were also instructed to return the affect product to Biome
DistributionWorldwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.