Recall Z-2235-2008 ยท JAK System, X-Ray, Tomography, Computed

Philips And Neusoft Medical Systems Co., Ltd.

NeuViz Dual series computed Tomography Scanner System, X-Ray System The CT scanner is a whole body x-ray computed tomography scanner which features a continuously rotating tube-detector system and functions according to the fan beam principle.

DateSep 17, 2008 (initiated Dec 17, 2007)
StatusTerminated
ReasonLabeling of patient position on scanned image does not match actual patient orientation.
Root causeSoftware design
ActionOn 12/17/08 the distributor sent a letter dated 12/13/07 to their customers explaining the problem and under what circumstances it can occur, the actions that should be taken by the customer/user in order to prevent risks for patients or users, and that the customer/user will be contacted for scheduling of a software upgrade by the distributor. Contact NEUISYS Service Support Department at 1-877-299-9052 for assistance.
Quantity15 units
DistributionDistribution in US: CO, FL, KY, NC, SC, TN, and TX.