Recall Z-0007-2009 ยท JIY Photometric Method, Iron (Non-Heme)
Roche Diagnostics Corp.
Roche Diagnostics IRON2, Iron Gen. 2, COBAS INTEGRA, cobas c systems, Roche Diagnostics, Indianapolis, IN.; 03183696122. An in-vitro diagnostic reagent system intended for use on COBAS Integra systems for the quantitative determination of iron in serum and plasma.
| Date | Oct 9, 2008 (initiated Jun 11, 2008) |
| Status | Terminated |
| Reason | Some of the labels are glossy, resulting in the bar code being difficult to read by the bar code reader. |
| Root cause | Labeling Change Control |
| Action | Consignees were notified via an Urgent Medical Device correction letter dated 6/11/08 informing users of the problem and providing them with work-around instructions. The instructions suggests 1) inserting the cassette rack slowly; 2) using mainly reagent slots A, B, or C; 3) tilting the rack slightly clockwise so the barcode scanner does not read the barcode at an exact 90 degree angle; and 4) confirming the rack is properly seated on the track and displayed on the Cassette Status screen. If the cassette barcode does not register after following the Work-Around instructions, then customers |
| Quantity | 523 |
| Distribution | Nationwide Distribution |
| FDA record | Open on fda.gov |