Recall Z-2195-2008 ยท IYE Accelerator, Linear, Medical

Varian Medical Systems Oncology Systems

Varian Medical Systems FastPlan, versions 5.5 and 5.5.1, model number HZ3, manufactured by Varian Medical Systems, Palo Alto, CA The product is a radiation treatment planning system.

DateSep 16, 2008 (initiated May 29, 2008)
StatusTerminated
ReasonDue to issues with image orientation verification, a patient may receive all or some of prescribed high dose to healthy tissue instead of intended area.
Root causeLabeling False and Misleading
ActionRecall was initiated by the firm on 05/29/2008. A product notification letter /field safety letter was distributed to all affected users, with a description of the problem and user corrective action steps. In addition, the letter will also be distributed to Varian Sales, Marketing and Service Organizations. Letters were sent to consignees via certified mail with return receipt mailing to US consignees. For foreign consignees, letters will be translated as necessary and distributed with return proof of notification. Varian Medical can be contacted at 1-650-483-3153 for assistance.
Quantity182
DistributionProduct was distributed throughout the US and to Australia, Brazil, Chile, France, Germany, Hong Kong, India, Israel, Japan, China, Malaysia, Russian Federation, Singapore and Spain.