ClearanceBrief
Recall Z-2173-2008 ยท LKK Pump, Infusion, Implanted, Programmable

Medtronic Neuromodulation

Model 8709SC Indura 1P One-Piece Intrathecal Catheter Medtronic Inc., Minneapolis, MN 55432 A part of the SynchroMed II Infusion system, which is designed to restore and provide intrathecal administration of drug therapy.

DateAug 27, 2008 (initiated Apr 14, 2008)
StatusTerminated
ReasonThe SC catheters cannot completely engage with the Model 700-04M portal connector. The result of this condition is a possible leakage or disconnect of the SC catheter from the portal during the continuous therapy trial period.
Root causeDevice Design
ActionMedtronic informed all Belgium implanting and managing physicians of the SynchroMed II infusion system. Each customer was mailed a Dear Healthcare Professional letter (Medtronic Urgent: Field Safety Notice) on 4/3/08 providing important safety information and patient management recommendations regarding the device incompatibility. The firm recommends that healthcare professionals do not use SC catheters if a connection with an implantable access port is required. Also, if a continuous infusion trial is required with use of an implantable access port, they should use a model 8711 intrathecal
Quantity232 total from all models
DistributionInternational Distribution --- country of Belgium.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.