Recall Z-2172-2008 ยท LKK Pump, Infusion, Implanted, Programmable
Medtronic Neuromodulation
Model 8596SC Intrathecal Catheter Pump Segment Revision Kit Medtronic Inc., Minneapolis, MN 55432 A part of the SynchroMed II Infusion system, which is designed to restore and provide intrathecal administration of drug therapy.
| Date | Aug 27, 2008 (initiated Apr 14, 2008) |
| Status | Terminated |
| Reason | The SC catheters cannot completely engage with the Model 700-04M portal connector. The result of this condition is a possible leakage or disconnect of the SC catheter from the portal during the continuous therapy trial period. |
| Root cause | Device Design |
| Action | Medtronic informed all Belgium implanting and managing physicians of the SynchroMed II infusion system. Each customer was mailed a Dear Healthcare Professional letter (Medtronic Urgent: Field Safety Notice) on 4/3/08 providing important safety information and patient management recommendations regarding the device incompatibility. The firm recommends that healthcare professionals do not use SC catheters if a connection with an implantable access port is required. Also, if a continuous infusion trial is required with use of an implantable access port, they should use a model 8711 intrathecal |
| Quantity | 232 total from all models |
| Distribution | International Distribution --- country of Belgium. |
| FDA record | Open on fda.gov |