Recall Z-1879-2008 ยท KRO Prosthesis, Knee, Femorotibial, Constrained, Cemented, Metal/Polymer
Biomet, Inc.
Biomet OSS Interlok IM Stem w/Screw, straight collar, 11 mm x 225 mm, TI6AL4V alloy, sterile, Biomet Orthopedics, Inc., Warsaw, IN; REF 150490. Orthopedic implant intended as part of a replacement knee joint.
| Date | Sep 11, 2008 (initiated May 15, 2008) |
| Status | Terminated |
| Reason | Mislabeled--The product is not collared, although the label states that it is collared. A non-collared component will provide a reduced contact with bone interface and/or a delay in the procedure may be incurred while clarifying package contents. |
| Root cause | Labeling False and Misleading |
| Action | On May 15, 2008 an Urgent Medical Device Removal Notice was issued to consignees and implanting physicians notifying them to return the product and of risks involved, respectively. Any adverse reactions which may be experienced with the use of this product should be reported to FDA's MedWatch Program by phone at 1-800-FDA-1088, by Fax at 1-800-FDA-0178, by mail at MedWatch, HF-2, FDA, 5600 Fishers Lane, Rockville, MD 20852, or the MedWatch website at www.fda.gov/medwatch. If you have any questions contact Mary Hardesty, Manager, Corrective Actions, Biomet Orthopedics, Inc. at 1-800-348-9500. |
| Quantity | 14 for all products. |
| Distribution | Class II Recall - Nationwide Distribution --- including states of California, Indiana, Kentucky, Massachusetts and Virginia. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K002757 | ONCOLOGY SALVAGE SYSTEM Biomet, Inc., Nov 24, 2000 |