Recall Z-2180-2008 ยท LLZ System, Image Processing, Radiological

AGFA Corp.

Impax 6.x Clients including: 6.3 all versions below 6.3.1 SU5, 6.2 all versions below 6.2.1 SU14 and 6.0 all versions. The Impax Systems are Picture Archiving and Communications system (PACs). PAC systems are used in the acceptance, transfer, display, storage and digital processing of medical images.

DateSep 4, 2008 (initiated Jun 4, 2008)
StatusTerminated
ReasonThe IMPAX client failed to "refresh" the image area upon receiving a study retrieved from archive. The system also erroneously displays this study retrieved from archive as the "active" study.
Root causeSoftware design
ActionAn "Urgent Safety Notice" letter was sent to all the sites using potentially affected units on/about June 4, 2008. Dealers were notified via letter on June 9, 2008. The letter requests that all users be reminded that Agfa recommends the system is configured so that demographic information is not removed from the image viewing area. They will be provided with the appropriate Service Update, based on the current version of IMPAX Client 6.x in use at their site. They were advised to distribute the information within their facility to all those who need to be aware of it. A FAX Back form was reque
Quantity371 units
DistributionU.S., Canada, Europe, Asia Pacific and Latin America.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K022292IMPAX DIAGNOSTIC DISPLAY STATION,IMPAX CLINICAL REVIEW STATION,IMPAX FOR CARDIOLOGY CLINICAL REVIEW SATATION,IMPAX ORTHO