Recall Z-2178-2008 ยท MQH Microtools, Assisted Reproduction (Pipettes)

Cook Vascular Inc.

Flexipet Manipulation Pipette 300 micron. The product is boxed with 1 vial of 10 pipettes. This product is used for the intracytoplasmic single sperm injection of oocytes.

DateSep 16, 2008 (initiated Dec 17, 2007)
StatusTerminated
Reasonmislabeled - 300 micron flexipets were labeled as 80 microns
Root causeEmployee error
ActionThe sales representatives were notified via email on 12/14/07 of the problem. The recalling firm issued recall letters on 12/18/07 via overnight mail. The letter informed the customers of the problem and the need to return any product. The letter also instructed customers to notify any sub accounts. Contact Cook Vascular at 1-724-845-8621 for assistance.
Quantity35 vials
DistributionThe product was shipped to a medical facility in TN and sales reps in NJ and MD. The product was also shipped to distributors in Indonesia, Australia, and Ireland
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K983596INTRACYTOPLASMIC SPERM INJECTION (ICSI) MICRO-INJECTION PIPETTES, HOLDING PIPETTES, DENUDING PIPETTES, ASSISTED HATCHING