Recall Z-1902-2008 ยท KFM Pump, Blood, Cardiopulmonary Bypass, Non-Roller Type

Levitronix, Llc

Levitronix CentriMag Back-Up Console Article Number: 201-30200; extracorporeal blood circulation system.

DateJul 23, 2008 (initiated Mar 17, 2008)
StatusTerminated
ReasonStops pumping: Interruption of CentriMag System support may occur when using a Valleylab Force FX-C Electrocautery Unit
Root causeDevice Design
ActionLevitronix issued an Urgent - Voluntary Device Correction letter on 3/17/08 mailed to the US distributor and requested to contact their accounts. The letter revises the warning in the CentriMag Console Manual. Levitronix is issuing revised manuals. On March 13, 2008 the firm emailed and mailed (FedEx) to the one Canadian Hospital and Health Canada was informed of the Voluntary Advisory Notice via email on Wednesday, March 19, 2008. Rest of the World: Wednesday, March 12, 2008 through Monday, March 17, 2008. The Notice was issued as a Voluntary Advisory Notice after review and follow up wi
Quantity237 units
DistributionWorldwide Distribution: USA, Canada, Switzerland, UK, Italy, France, Greece, Spain, Slovenia, Czech, Turkey, South Africa, Argentina, Chile, Panama, Singapore, Taiwan, Israel, and Saudi Arabia.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K051209LEVITRONIX CENTRIMAG BACK-UP CONSOLE
Levitronix, LLC, Sep 8, 2005