Recall Z-2352-2008 ยท FKP System, Dialysate Delivery, Single Patient

Hach Co

HACH SteriChek Sensitive Total Chloramines and Residual Chlorine Reagent Strips, Tests for feed water and rinse water down to 0.1 ppm (mg/l), packaged in bottles containing 100 strips each, 5 bottles per kit, Part #811909. The firm name on the label is Hach, Ames, IA. The product is used in measuring for low levels of total chlorine (including chloramines and free chlorine) in feed water used to

DateSep 21, 2008 (initiated May 23, 2008)
StatusTerminated
ReasonInconsistencies in total chlorine and free chlorine levels may result in providing inaccurate false positive or false negative results
Root causeNonconforming Material/Component
ActionThe firm initiated phone calls on 5/22/08 and explained the reason for recall and requested the product be quarantined. The customer was informed they would receive a recall letter providing instructions for return, replacement, or credit. During the phone call, some of the customers asked if they should recall to the end user level and they were told to do so, but this instruction was not verbally given to everyone. The recalling firm issued a letter dated 5/23/08 via overnight mail which explained the reason for recall and requested return of the product. The letter requests the custo
Quantity80,110 bottles
DistributionNationwide and Foreign distribution to Canada and Denmark.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K003615STERICHEK SENSITIVE-FEED WATER AND RINSE WATER REAGENT STRIPS
Environmental Test Systems, Inc., Feb 1, 2001