Recall Z-2160-2008 ยท IZJ Camera, X-Ray, Fluorographic, Cine Or Spot

Philips Medical Systems North America Co. Phillips

General X-Ray System types with Digital Spot Imaging (DSI) software

DateAug 25, 2008 (initiated May 23, 2008)
StatusTerminated
ReasonX-ray images may be stored in the wrong patient file or corrupted, images from two different patients may be combined in one patient's folder or may be combined on one image (corrupted).
Root causeSoftware design
ActionOn 5/20/08 the firm issued an URGENT - DEVICE CORRECTION letter to the consignees advising of the problem, requesting consignees to check the stored or printed images to ensure that it represents the data and images captured from the examination. The firm will issue a Field Change Order to their service representatives providing new software, when software is available. This should be issued by the end of June.
Quantity41 units
DistributionNationwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K920793PHILIPS DSI SYSTEM RELEASE 3.2
Philips Medical Systems, Inc., Jul 17, 1992