Recall Z-2155-2008 ยท HWC Screw, Fixation, Bone

Acumed LLC

3.5mm X 32.5mm Cortical Screw used with the Polarus Humeral Fixation System,. Product labeled in part, "Polarus Humeral Fixation System& SIZE: 3.5mm X 32.5mm Cortical Screw... REF HCO3325-S... Acumed 5885 NW Cornelius Pass Road Hillsboro, OR 97124-9432 ". Polarus Humeral Fixation System, 3.5mm X 32.5mm Cortical Screw are packaged in a sterile inner package which are placed into an outer box.

DateAug 28, 2008 (initiated Dec 27, 2005)
StatusTerminated
ReasonPackage of 3.5mm X 32.5mm cortical screw may contain 3.5mm X 22.5mm cortical screw.
Root causeProcess change control
ActionThe firm issued a Product Removal notification on 12-27-2005 to their customers. The notification was sent by either FAX or by e-mail. The Product Removal Notification informed their customers of the issue, asked them to remove the screws from inventory and return the screws to Acumed. A reminder notification, dated 1/20/08 was issued to customers identifying the product and the problem and requesting the product be returned.
Quantity112 pieces
DistributionWorldwide Distribution --- USA including states of CA, KY, FL, TX, OH, OR, NY, ID NJ, NE, LA, WI, UT, MO, MS, GA, AZ and countries of United Kingdom, Italy, Puerto Rico, Turkey, Spain, Japan, South Africa, and Belgium.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K942340CORTICAL BONE SCREW
Acu Med, Inc., Dec 5, 1994