Recall Z-2119-2008 ยท HWC Screw, Fixation, Bone
Acumed LLC
10mm Bone Graft Drill. Product labeled in part, "ACUMED 5885 NW Cornelius pass Road Hillsboro, OR. 97124-9432... Size: 10mm Bone Graft Drill... REF BG-8020-S ... LOT W54076".
| Date | Sep 20, 2008 (initiated Feb 14, 2008) |
| Status | Terminated |
| Reason | Product sterility may be compromised. |
| Root cause | Packaging process control |
| Action | The firm issued four separate Urgent Notification: Product Removal were issued by the firm on 2-14-2008, 2-20-2008, 2-26-2008, and 3-25-2008 for the issue of poorly sealed sterile pouches. The notifications were reportedly sent by either FAX or e-mail. The firm explained the issue and requested customers to return the specified products. On 7/16/08, the firm issued a second recall notification to their customers. This letter informed the customer of the recall, listed the products affected, stated the reason for the recall, tells customers, "return existing inventories of the recalled produ |
| Quantity | 31 pcs |
| Distribution | Worldwide Distribution ---- USA including states of WA, NE, NC, UT, FL, CA, OR, TX, NY, IL, OH, MT, and Puerto Rico, and countries of United Kingdom, Australia, and Greece. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K960711 | ACUMED BONE GRAFT SYSTEM Acu Med, Inc., May 7, 1996 |