ClearanceBrief
Recall Z-2068-2008 ยท MNI Orthosis, Spinal Pedicle Fixation

Allez Spine, LLC

Allez Spine Laguna Pedicle Screw System, Non-Conical Screw Size (diameter x length) 7 x 40mm, Non-winged, Catalog # LG-PR0740

DateSep 17, 2008 (initiated Nov 7, 2007)
StatusTerminated
ReasonThis recall was initiated after Internal Engineering Testing of the conical seat design pedicle screws of sizes 5, 6 and 7 showed a potential of partial displacement of screw shank when used in combination with the single piece locking nuts. This partial displacement of the screw shank could finally lead to total separation and failure of the construct.
Root causeDevice Design
ActionAllez Spine informed all parties who were in possession of the affected pedicle screws at the time of the initiation of this field action by phone on 11/08/2007 and requested the return of their inventory. All customers and consignees were contacted by the firm.
Quantity9
DistributionNationwide to AZ, CA, CO & UT
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K050060LAGUNA PEDICLE SCREW SYSTEM
Allez Spine, LLC, May 4, 2005
K051275LAGUNA MONOAXIAL PEDICLE SCREW SYSTEM
Allez Spine, LLC, Dec 9, 2005