ClearanceBrief
Recall Z-2103-2008 ยท IYE Accelerator, Linear, Medical

Siemens Medical Solutions USA, Inc

Radiation therapy system - Coherence Therapist System 2.1 equipped with Coherence Therapist RTT2.1 or Primeview 3i R2.1, part number 8147667. Medical charged particle Radiation therapy system, part of the firm's beam limiting device/accessory. The product provides data processing, and the SYNGO standard medical imaging platform to provide comprehensive oncology workflow solutions for radiation t

DateSep 17, 2008 (initiated May 15, 2008)
StatusTerminated
ReasonSoftware - Under certain circumstances, the portal image may be overwritten, which may lead to incorrect dosage.
Root causeSoftware design
ActionAn Urgent Medical Device Correction - Customer Safety Advisory Notice letter (TH014/08/S) was released on May 15, 2008. A new software patch is under construction and is planned for release in June 2008. If you have any questions contact Roland Betz at (925) 246-8407.
DistributionClass II Recall - Product distribution information pending from firm.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K060226ONCOR EXPRESSION
Siemens Medical Solutions USA, Inc., Mar 15, 2006