Recall Z-2106-2008 ยท DRA Catheter, Steerable

Hansen Medical Inc

Artisan Control Catheter, model number 04454, manufactured by Hansen Medical, Inc., Mountain View, CA. Steerable guide catheter for human use intended to facilitate manipulation, positioning and control for collecting electrophysiological data within the heart atria.

DateSep 11, 2008 (initiated May 21, 2008)
StatusTerminated
ReasonThe catheter's flexible bellows portion may develop a leak. This has the potential to cause loss of hemostasis, flush fluid leakage, and/or introduction of air into the catheter with a risk of subsequent embolism.
Root causeDevice Design
ActionOn May 21, 2008 an Urgent: Voluntary Product Recall Notification was issued. All customers who received affected product were verbally contacted directly by the firm, and were sent a follow up letter, asking them to segregate the product. Replacements will be shipped to them. If you have questions please contact Doug Worth at 1-650-404-5940.
Quantity369 units
DistributionClass II Recall - Worldwide Distribution ---- including USA and countries of UK, Italy, Germany, and Czech Republic.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K052480HANSEN MEDICAL CATHETER CONTROL SYSTEM (CCS) AND ACCESSORIES, HANSEN MEDICAL STEERABLE GUIDE CATHETER (SGC) AND SHEATH