Recall Z-2231-2008 ยท CGZ Electrode, Ion-Specific, Chloride

Abbott Laboratories, Inc

ARCHITECT c8000 Processing Module, for in vitro diagnostics; List Number 1G06-01 and 1G06-11, Affected software: v2.20, v2.20DB, v2.60, v3.10, 3.11 and 3.12; distributed by Abbott Laboratories, Irving, TX. The Abbott ARCHITECT cSystem is designed to perform automated, random access, clinical chemistry analyzer which utilizes spectrophotometry (monochromatic and bichromatic modes of measurement)

DateSep 11, 2008 (initiated May 23, 2008)
StatusTerminated
ReasonSoftware defect can allow tests ordered for one sample to be aspirated from a different sample, reporting erroneous results for the affected sample.
Root causeSoftware design
ActionFirm notified all ARCHITECT c8000 customers with a correction letter and response form distributed on 5/23/2008. The letter provided actions the users could take until an updated software version was available. For assistance, contact Abbott at 1-877-4ABBOTT.
Quantity2,378 units worldwide (431 in the USA; plus 1,947 International)
DistributionWorldwide distribution, including: USA (including Puerto Rico), Argentina, Australia, Austria, Bahamas, Belgium, Bermuda, Bosnia and Herzegovina, Brazil, Cayman Islands, Canada, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Dominican Republic, Ecuador, El Salvador, Finland, F
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K980367AEROSET SYSTEM
Abbott Laboratories, Apr 1, 1998