ClearanceBrief
Recall Z-0090-2009 ยท IYE Accelerator, Linear, Medical

Siemens Medical Solutions USA, Inc

Coherence Therapist System Part number 7339125 Medical charged-particle radiation therapy system.

DateOct 16, 2008 (initiated Nov 6, 2007)
StatusTerminated
ReasonFlat panel positioning calibration could be off by as much as 4 mm without the machine discovering detail.
Root causeSoftware design
ActionService calibration inspection/correction Update instructions TH038/07/S was issued on 11/06/2007 to all customers. The instructions identified the affected products, stated the reason for the update, and listed the steps to update the software. Customer notification was also issued. Software patch planned for release in June 2008.
Quantity1
DistributionWorldwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K060226ONCOR EXPRESSION
Siemens Medical Solutions USA, Inc., Mar 15, 2006