Recall Z-2090-2008 ยท LTJ Prostate-Specific Antigen (Psa) For Management Of Prostate Cancers

Siemens Medical Solutions Diagnostics

COAT-A-COUNT PSA IRMA kit (IKPS1, 2, 5, X). The device is intended as an aid in monitoring patients for disease progress or response to therapy.

DateSep 16, 2008 (initiated May 21, 2008)
StatusTerminated
ReasonThe field action was initiated after the firm confirmed that the Coat-A-Count PSA IRMA kit (IKPS1, 2, 5, X) exhibits a low bias, which is evident when comparing results with other methods, including the IMMULITE PSA kits. Coat-A-Count PSA IRMA results are, for example, approximately 20% lower when compared to results obtained with the IMMULITE 2000 Third Generation PSA assay (L2KUP).
Root causeOther
ActionA Customer Bulletin (Urgent Field Safety Notice) was prepared and sent by fax or mail to all affected Siemens Healthcare Diagnostics Branches on May 21, 2008 both in the United States and elsewhere for communication with the affected customers. This bulletin advises the customer of the problem and the discontinuation of the product. Customers were also instructed to forward this notification to anyone to whom you may have distributed this product and if they have additional questions, please contact your local technical service center @ 1-877-229-3711. A Confirmation (Fax-back) Form is incl
Quantity64 kits
DistributionNationwide to: CA, FL, IL, IN, MD, MI ; Worldwide to: Bolivia, Chile, Egypt, Cyprus, Korea, Philippines, Taiwan, Uruguay