Recall Z-2152-2008 ยท GIO Tube, Collection, Capillary Blood

Opti Medical Systems Inc

Vitrex Blood Gas Capillary Tubes, MC0024, REF 189301, 250 pcs, 1,45 x 1,85 x 125 mm app. 200l, Heparinized: 6 IU Na & 9 IU Li/100 l,

DateAug 14, 2008 (initiated May 8, 2008)
StatusTerminated
ReasonThis lot of Heparinized Capillary Tubes contain contaminated lithium heparin.
Root causeNonconforming Material/Component
ActionOpti Medical sent a Recall Notice, dated May 8, 2008 to all consignees. Consignees were notified via email on/about May 5th thru 7th , 2008. They were instructed to discontinue use of the affected lot and to immediately return all remaining product to OPTI Medical for replacement.
Quantity31750 tubes
DistributionWorldwide Distribution: USA, Puerto Rico, Austria, China, Chile, Cyprus, Great Britain, Japan, Netherlands, and Senegal