Recall Z-2233-2008 ยท JQW Station, Pipetting And Diluting, For Clinical Use

Hamilton Co

Hamilton Soft Grip Pipettes, 1 ml adjustable volume, model number 55019-40 Product is indicated for manual pipetting.

DateSep 17, 2008 (initiated May 7, 2008)
StatusTerminated
ReasonManufacturing issue with spring may restrict product to drawing less than 1/2 of its stated measured capacity
Root causeDevice Design
ActionRecall initiated May 7 2008. Notification issued via certified overnight mail explains the problem and instructs customers to inventory the affected pipettes. The firm will contact its customers and supply replacement lower-body assemblies, repair instructions and return information so that affected parts can be returned to Hamilton. A fax-back form is supplied with each notification. Responses will be tracked. Contact Hamilton Co. at 1-775-558-3000 for assistance.
Quantity305 units
DistributionProduct distributed throughout the US and to Puerto Rico, New Zealand, Australia, Japan and Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.