Recall Z-2081-2008 ยท JAQ System, Applicator, Radionuclide, Remote-Controlled

Varian Medical Systems, Inc.

VariSource CT/MR Ring & Tandem Applicator Set, catalog # AL13017000, containing 32mm titanium ring applicators, each identified as part numbers AL07362000, AL07363000, and AL07364000.

DateSep 16, 2008 (initiated Mar 12, 2008)
StatusTerminated
ReasonMedical device may exhibit inaccurate positioning and lead to unintended dose delivery during brachytherapy treatment.
Root causeDevice Design
ActionThe firm notified medical device users by letter on 03/12/08. The notification provided updated instructions for use for the CT/MR Ring and Tandem Combination Applicator Set. Users were advised to characterize all ring applicators to develop an offset correction for each dwell position in order to construct a plan to deliver the dwell positions accurately and reliably. The characterization can be done by creating a set of autoradiographs with intended dwell positions and define treatment plan shifts in order to deliver dwell positions in intended locations. Users were advised that a subsequent
Quantity120 units
DistributionThe medical devices were distributed to radiation oncology facilities nationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K030110CT HDR TANDEM/RING APPLICATOR WITH RECTAL RETRACTOR
Mick Radio-Nuclear Instruments, Inc., Apr 3, 2003