ClearanceBrief
Recall Z-1920-2008 ยท DTZ Oxygenator, Cardiopulmonary Bypass

Medtronic Perfusion Systems

Medtronic Extracorporeal Circuit with bio-active surface; Cardiopulmonary Bypass Oxygenator, Catalog Numbers: CB0B90R3, CB0L44R15, CB0X97R10, CB1V99R2, CB1Y06R14, CB1Z42R15, CB1Z69R12, CB1Z69R12, CB1Z85R16, CB2503R1, CB2K59R8, CB2K59R8, CB2S52R4, CB2U05R1, CB2W86R2, CB3E29R5, CB3L87R4, CB3U53R7, CB3X44R4, CB4741R15, CB4E05R6, CB4F32R3, CB4F32R3, CB4F32R3, CB4G10R5, CB4Q81R, CB4S04R2, CB4T1

DateJul 24, 2008 (initiated May 9, 2008)
StatusTerminated
ReasonMedtronic is recalling selected Carmeda-coated cardiopulmonary bypass (CPB) products (e.g., oxygenators, reservoirs, pumps, cannulae and tubing packs) that were manufactured with contaminated heparin. Medtronic's heparin supplier recently reported that it had provided some batches to Medtronic that were contaminated with oversulfated chondroitin sulfate (OSCS).
Root causeNonconforming Material/Component
ActionMedtronic issued a Press Release on 05/07/2008. An Urgent Medical Device Recall letter was sent to customers on May 07, 2008. The letter identifies affected devices, explains the situation and requests that the impacted product be returned to Medtronic for disposal. The end user will be requested to complete a Recall Certificate acknowledging the receipt of this information.
Quantity311 US, 27 OUS
DistributionWorldwide Distribution: USA, Australia, Belgium, Canada, Czech Republic, France, Italy, Japan, Norway, and Taiwan.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K891687EXTRACORPOREAL CIRCUIT WITH BIO-ACTIVE SURFACE
Medtronic Blood Systems, Inc., May 31, 1989