Recall Z-0207-2009 ยท JQP Calculator/Data Processing Module, For Clinical Use

Sunquest Information Systems, Inc.

Sunquest Laboratory versions 5.3, 5.4, 6.1, 6.2 and 6.3. Sunquest Laboratory Information System is comprised of several modules which are involved in aspects such as blood banking, collection management, specimen tracking, and use with microbiology and molecular diagnostics. Key capabilities include: 1) Autoverify results directly from analyzers and reference laboratories 2) Track tubes and cont

DateOct 31, 2008 (initiated Apr 17, 2008)
StatusTerminated
ReasonSunquest has identified a problem with IGO interfaces when performing an Online File Cleanup (OFC) that can lead to the test results of one patient filing to another patient when the order codes are the same. Performing an OFC prior to all results being assigned an accession number may cause the patient results to remain in a temporary file where they can be incorrectly associated to the wrong pat
Root causeComponent design/selection
ActionCustomers were sent a preliminary notice on 4/17/2008 followed by customer notice PSN-08-04 on 05/07/2008 via email. The notice advises that this notice is follow-up to the Preliminary notice distributed on 4/17/08 and applies to all clients who have an Instrument Generated Order (IGO) interface. Sunquest has identified a problem with IGO interfaces when performing an Online File Cleanup (OFC) that can lead to the test results of one patient filing to another patient when the order codes are the same. Performing an OFC prior to all results being assigned an accession number may cause the pati
Quantity419 clients had the affected code
DistributionWorldwide Distribution --- including USA and countries of Canada, Bermuda, Ireland, United Kingdom, United Arab Emirates, and Saudi Arabia.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.