Recall Z-2059-2008 ยท DIP Enzyme Immunoassay, Diphenylhydantoin

Abbott Laboratories Inc.

ABBOTT Clinical Chemistry Phenytoin, List Number: 1E07-20, is used for the quantitation of phenytoin in human serum or plasma.

DateSep 16, 2008 (initiated Feb 15, 2008)
StatusTerminated
ReasonAn Abbott investigation has determined that the phenytoin reagent does not maintain an onboard stability of 28 days as stated in the phenytoin package insert. In addition, the 14 day calibration interval is also not being met. Further investigation is required to determine the root cause of this issue. Until the cause has been determined and a corrective action defined, specific instructions ar
Root causeOther
ActionOn February 15, 2008 A Product Correction letter with instrument specific attachments was provided to all Clinical Chemistry Phenytoin customers that have received the affected lots. A customer reply form was included in all US customers letter. The letter instructed customers to follow the attached instrument specific actions as described in the attachments (Attachment A: ARCHITECT cSYSTEM, Attachment B: AEROSET). Customers were also instructed to forward a copy of the Product Correction letter to any other laboratories to which they may have forwarded the affected product. If you have ques
Quantity3266 Worldwide, 2009 Domestically
DistributionNationwide and Worldwide to: Australia, Canada, Chile, Germany, Hong Kong, New Zealand, Panama, Singapore, Uruguay, Singapore
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K993026ABBOT AEROSET PHENYTOIN ASSAY & ABBOT AEROSET PHENYTOIN CALIBRATORS
Syva Co., Jan 27, 2000