Recall Z-1862-2008 ยท IYE Accelerator, Linear, Medical

Accuray Inc

Accuracy Cyberknife Robotic Radiosurgery System, medical charged particle radiation therapy device. Model number 020700 (axum/standard treatment couch) and 021756 hand controller. The device is indicated for treatment planning and image-guided stereotactic radiosurgery and precision radiotherapy of lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.

DateSep 16, 2008 (initiated May 1, 2008)
StatusTerminated
ReasonCouch may move unexpectedly, which may result in patient impacting the linear accelerator.
Root causeSoftware design
ActionAll affected sites were notified via customer advisory notice. The Standard Treatment couch firmware will be updated and all affected sites will be retrofitted. Contact Customer Service at 1-877-668-8667 (USA) or 1-408-716-4700 (non-USA) or customersupport@accuray.com for additional information.
Quantity105 units
DistributionProduct was distributed throughout the US and to China, Taiwan, Korea, Italy, Malaysia, Germany, Netherlands, Vietnam, Greece, Spain, and Puerto Rico.