ClearanceBrief
Recall Z-2013-2008 ยท IYN System, Imaging, Pulsed Doppler, Ultrasonic

Philips Medical Systems

Biopsy Guide Starter Kits C5-2 used with the Philips HD3 Ultrasound System intended for diagnostic ultrasound imaging or fluid analysis of the human body.

DateSep 16, 2008 (initiated May 29, 2008)
StatusTerminated
ReasonHD3 Biopsy Guide Kits do not contain adequate labeling instructions for safe and effective use.
Root causeLabeling design
ActionOn 6/4/08, the firm sent letters to their customers via UPS. The letter informed the customers of the recall, and informed them of the following: "Customers who previously purchased and received a model C7-3 Biopsy Guide Starter Kit between November 29, 2005 and April 3, 2008 will receive a new kit at no charge, to replace the incomplete C7-3 kit previously purchased. New C7-3 kits will be distributed to affected customers by June 30, 2008. Please discard immediately the old C7-3 Biopsy Guide Kit... Customers who previously purchased and received model C5-2 and/or model L9-5 will find the res
Quantity79 units
DistributionProducts are distributed to 5 medical facilities located in NY, WA, GA, VA and MI and to 51 foreign consignees located in AR, AT, CH, CN, CO, CY, DE, DK, ES, FR, GB, GR, HK, HR, HU, IN, IT, KR, LB, LV, MA, MD, MY, NI, NL, PL, SG, SV, TH, TR, TW, ZA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K031552MYCOLOR 202 DIAGNOSTIC ULTRASOUND SYSTEM AND TRANSDUCERS
Medison America, Inc., May 30, 2003