Recall Z-1819-2008 ยท MNI Orthosis, Spinal Pedicle Fixation

Allez Spine, LLC

Allez Spine Laguna Pedicle Screw System: IT-FN1001, Laguna Sterilization Tray # 1 (Part Number: 54003); IT-FN1002, Laguna Sterilization Tray # 2 (Part Number: 54004)

DateSep 2, 2008 (initiated Apr 14, 2008)
StatusTerminated
ReasonThis field action was initiated because the cleaning, decontamination and sterilization procedures specified in the "Instructions for Use" for the impacted systems are incorrect. The use of these parameters in performing cleaning, decontamination and sterilization procedures may not result in clean and sterile product.
Root causeError in labeling
ActionAllez Spine issued a safety notice with attached fax-back attachment on April 14, 2008 via FedEx to all its hospital users, distributors and field representatives informing them of the failure of the sterilization validation and requesting them to use the new recommended sterilization parameters. The notice included copies of the updated Instructions for Use for the Allez Spine Laguna Pedicle Screw System and the Allez Spine Del Mar Pedicle Screw System. The safety notice requested that the revised recommended instructions be used for all future cleaning, decontamination and sterilization. Cop
Quantity64 systems
DistributionNationwide to AZ, CA, CO, GA, IL, NV, TN, TX, UT
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K050060LAGUNA PEDICLE SCREW SYSTEM
Allez Spine, LLC, May 4, 2005