Recall Z-2056-2008 ยท IYE Accelerator, Linear, Medical

Accuray Inc

CyberKnife Robotic Radiosurgery System radiation therapy device with Multiplan treatment Planning Software versions 1.5.2 and higher.

DateSep 17, 2008 (initiated Apr 9, 2008)
StatusTerminated
ReasonSample beam data (which should not be used to treat patients) may differ from actual radiation output of an installed product, and which may be used by users.
Root causeSoftware design
ActionOn April 16, 2008, the firm issued notifications via fedex to affected sites. In addition, installation and upgrade procedures will be updated to incorporate a step which decommissions sample beam data. If you have questions, contact Accuray Customer Support at 1-877-668-8667 (USA) or 1-408-716-4700 (non-USA) or send and email to customersupport@accuray.com.
Quantity109 units
DistributionProduct was distributed to 104 medical centers throughout the US and to Vietnam, China, Korea, Taiwan, Malaysia, Turkey, France, Italy, Spain, Greece and Germany.