Recall Z-1811-2008 ยท NAY System, Surgical, Computer Controlled Instrument

Intuitive Surgical, Inc.

INTUITIVE SURGICAL DA VINCI SURGICAL SYSTEM, MODE; S Cardiac Probe Gasper Instrument, Model IS2000, Part Number 420215-02; Product is manufactured and distributed by Intuitive Surgical, Inc. Sunnyvale, CA . The IS2000 CPG is used during cardiac ablation procedures.

DateSep 17, 2008 (initiated Apr 18, 2008)
StatusTerminated
ReasonThe product has a software interface problem. When the product is connected to the da Vinci IS2000 System it will not allow the System to recognize the Instrument which makes the IS2000 CPG instrument nonfunctional at all sites. Risks associated are loss of operability of the instrument, delay in surgery, and loss of dexterity.
Root causeSoftware design
ActionOn April 18, 2008 an Urgent Device Recall letter was issued to customers. Consignees were notified via e-mail and were instructed to return the product. Customers were instructed to segregate the product in a secure area for Intuitive Surgical CSRs. If you have any questions contact Karen Uyesugi at (408) 523-2100.
Quantity11
DistributionClass II Recall - Nationwide Distribution --- including states of NC, OH, GA and WI.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K070947INTUITIVE SURGICAL DA VINCI SURGICAL SYSTEM AND ENDOSCOPIC INSTRUMENTS AND ENDOWRIST CARDIAC PROBE GRASPER
Intuitive Surgical, Inc., Feb 14, 2008