ClearanceBrief
Recall Z-1781-2008 ยท NSA Over-The-Counter Automated External Defibrillator

Philips Medical Systems

Philips "HeartStart Home" defibrillator, model M5068A and is sold in the US over the counter. The HeartStart is an automated external defibrillator for use on site and consists of a computer with digital interface and cartridge containing pads.

DateJul 9, 2008 (initiated Apr 30, 2008)
StatusTerminated
ReasonButtons Sticking - The On/Off and Shock buttons on the front panel of the device may stick in place when pressed and fail to respond to the button press, causing a delay in the delivery of a defibrillation shock when the system advises a shock.
Root causeComponent design/selection
ActionThree separate Urgent - Device Recall letter were sent out on April 30, 2008 via tractable letter using UPS. 1) The end-customer letter advises of the recall, contact information to arrange a replacement of the device and instructions for checking the equipment to insure it is working correctly. "If your HeartStart Home Defibrillator is one of the affected devices or if you have any questions about this recall, please contact Philips Healthcare at 1.800.263.3342, option 5, or visit www.HeartStartHome.com to arrange for a free replacement defibrillator. If you have given your HeartStart to so
Quantity389 units
DistributionWorldwide including USA and countries of Austria, Australia, Canada, Colombia, German, Denmark, Dominican Republic, Spain, France, England, Ireland, Italy, Japan, Netherlands, Norway, New Zealand, Panama, Sweden, Singapore, and South Africa.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K040904PHILIPS HEARTSTART HOME DEFIBRILLATOR, MODEL M5068A
Philips Medical Systems, Sep 16, 2004