Recall Z-1824-2008 ยท LNH System, Nuclear Magnetic Resonance Imaging

Resonance Technology Inc

Headset Model Number RTC-551-070 for use with the Serene Sound Digital MRI Compatible High Fidelity Stereo Sound System .

DateSep 2, 2008 (initiated May 14, 2007)
StatusTerminated
ReasonPotential safety concerns - Use of this product with a MRI may result in the headset's cord overheating and patients experiencing burns to the skin.
Root causeDevice Design
ActionThe strategy was notifying customers by written communication. The letter advises Resonance Technology, Inc. is voluntarily recalling the MRI Headset, Model No. RTC-2K and RTC- 551-070 because of potential safety concerns. Use of this product with a MRI may result in the headset's cord overheating and patients experiencing burns to skin. The letter also advises that Resonance Technology Inc. has designed and developed the significantly improved MRI Headset, Model No. RTC-661-060-09P-001 (upgrade) and began shipping this product in January 2007. Customers who received the Headset Model; RTC-
Quantity29 units
DistributionNationwide and Internationally: Canada, Japan, Taiwan, United Kingdom, Netherlands, Italy, Spain, Germany, Switzerland, France,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K994351M.R. VISION 2000 ULTRA AUDIO VISUAL SYSTEM / COMMANDER X6 SYSTEM
Resonance Technology, Inc., Mar 22, 2000