Recall Z-1698-2008 ยท GEH Unit, Cryosurgical, Accessories

Galil Medical

Urethral Warming System-Galil Medical Warming Catheter. The product is contained within prostate kits. This product is used in cryosurgery to prevent freezing of tissue.

DateSep 3, 2008 (initiated Dec 11, 2007)
StatusTerminated
ReasonThe warming catheter was manufactured with the inlet and outlet tubes switched.
Root causeProcess control
ActionThe recalling firm telephoned their customers in April 2008 to inform them of the problem and the need to return the product. There customers were also asked to contact the firm if they had actually used any of the kits with the serial numbers listed, even if they experienced no problem. As a follow up to the telephone call, a letter was issued to the customers. For assistance, please call 770-565-6166.
Quantity541 catheters
DistributionWorldwide Distribution --- USA including states of PA, SC, FL, NY, TN, and WA, and country of Canada