Recall Z-1794-2008 ยท FPK Tubing, Fluid Delivery

Mentor Texas, Inc

Y Tubing Set, 9'; 10/box; Part Number LAM-Y; distributed by Mentor Texas, LP, Irving, TX 75038.

DateSep 5, 2008 (initiated Apr 21, 2008)
StatusTerminated
ReasonDuring packaging integrity validation, it was found that various product packages could fail, compromising the sterility of the devices.
Root causeOther
ActionA Customer Letter was issued to all consignees by the firm on 4/21/2008. Consignees were informed that a recent testing of these products discovered a potential defect in the packaging configuration, which could result in a loss of package integrity. Consignees were asked to review inventory and dispose of product and contact firm for credit.
Quantity3,349 boxes.
DistributionClass II Recall - Worldwide Distribution ---- including USA, Puerto Rico and countries of Argentina, Austria, Bermuda, Brazil, Canada, China, Costa Rica, Cyprus, Denmark, England, Germany, Greece, Guatemala, India, Italy, Italy, Japan, Kuwait, Lebanon, Malta, Mexico, Morocco, Pakistan, Panama, Parag
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.