ClearanceBrief
Recall Z-1905-2008 ยท LDD Dc-Defibrillator, Low-Energy, (Including Paddles)

Physio Control, Inc.

Physio-Control LIFEPAK 20 defibrillator/monitor is indicated for the termination of certain potentially fatal arrhythmias, such as ventricular fibrillation and symptomatic ventricular tachycardia.

DateSep 16, 2008 (initiated Apr 16, 2008)
StatusTerminated
ReasonThe LIFEPAK 20 defibrillator / monitors with software versions 048, 052 and 054 have an increase in likelihood for an incorrect Shock Advisory Algorithm (SAS) decision if the Auto Analyze setting in AED mode is On, leading to incorrectly render of shock or no shock decision. When Auto Analyze is set to On in AED mode the device initiates the SAS analysis immediately (no waiting period or warning p
Root causeSoftware design
ActionOn April 16, 2008, a letter was sent to the firms customers, via certified mail explaining the recall, giving product and software version information, request for customer to forward the notification to all sites and recommending the following: "Keep the defibrillator in service... Check the Auto Analyze Setup Option for each of your devices. If Auto Analyze is set to ON, change the setting to one of the other two choices: OFF or After First Shock... Follow the enclosed instructions to confirm or change the Auto Analyze Setup Option setting...If you update your defibrillator with new softwa
Quantity21,102 units
DistributionProduct distributed nationwide and includes: 2 distributors, 157 direct accounts, 2,069 hospitals, 51 Department of Defense accounts, 58 miscellaneous government accounts, 3,295 first responders, 43 international accounts. International accounts are located in the following countries: ARGENTINA, A
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K063119LIFEPAK 20 DEFIBRILLATOR/ MONITOR
Medtronic Emergency Response Systems, Inc., Dec 22, 2006