Recall Z-1777-2008 ยท FNL Bed, Ac-Powered Adjustable Hospital

Hill-Rom, Inc.

Hill-Rom Century+ Bed Model P1400 HBSW siderail upgrade kits; accessory model numbers P9964-11, P9964-12, P9964-13, P-9964-14, P9964-15, P9964-16, P9964-41 and/or P9964-42.

DateAug 27, 2008 (initiated Apr 28, 2008)
StatusTerminated
ReasonA risk of entrapment will exist between the head and foot rails if installation instructions are followed.
Root causeOther
ActionHill-Rom issued an Urgent Medical Device Correction letter on 4/28/2008 notifying consignees that they will be visited by a qualified Hill-Rom service technician or designee and provide a revised copy of the instruction sheet. The technician will either assist the facility to ensure proper installation or verify by signature that the facility understands the issue and will verify their beds had the upgrade kits installed in such a way that an entrapment gap does not exist.
Quantity69
DistributionClass II Recall - Nationwide Distribution --- including states of Florida, Illinois, New Mexico, Texas and Virginia.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.