ClearanceBrief
Recall Z-1776-2008 ยท ITI Wheelchair, Powered

Pride Mobility Products Corp

Q-Logic Controller Software used on the following power wheelchairs: Q6000, Q6000XL, Q6000Z, R4000, Q600, Q600XL, Q610, Q1107, Q600E, and Dynamo 1107

DateAug 27, 2008 (initiated Apr 11, 2008)
StatusTerminated
ReasonSoftware - Watch dog timer feature was disabled.
Root causeSoftware design
ActionThe recalling firm issued an Urgent Device Correction letter on 4/11/08. This letter instructs the dealers to contact the end users so that the units can be upgraded to enable the watch dog feature.
Quantity4129
DistributionClass II Recall - Worldwide Distribution --- including USA and Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K011993QUANTUM BLAST
Pride Mobility Products Corp., Jul 13, 2001
K042612JAZZY 600
Pride Mobility Products Corp., Oct 6, 2004
K945936JAZZY
Pride Health Care, Inc., May 28, 1996