ClearanceBrief
Recall Z-1771-2008 ยท IOR Wheelchair, Mechanical

Medline Industries Inc

Bariatric Wheelchair, 26" Wide and 20" Deep Seat, 700 lb. Weight Capacity; Manufactured in Taiwan for Medline Industries, Inc., Mundelein, IL 60060 USA; Item #MDS809700 comes with Removable Full-Length Arms, Swing-Away Detachable Footrests, Hemi-Height Adjustable. item #MDS809750 comes with Removable Full-Length Arms, Swing-Away Detachable Elevating Legrests, Hemi-Height Adjustable.

DateAug 19, 2008 (initiated Mar 31, 2008)
StatusTerminated
ReasonInjuries to fingers. There is a potential for users to pinch their fingers between the seat bars when opening the wheelchairs. The injuries can be severe from fracture to severing.
Root causeComponent design/selection
ActionA Voluntary Field Correction letter dated 3/31/08 was issued to customers. The letter informed them of reports of users catching their fingers under the seat bars, which are identified on the chair with warning stickers. The customers were requested to change the backrest with the enclosed replacement upholstery kit, properly dispose of the old back rest, and place the enclosed cautionary tags on both sides of the chair seat. Instructions for installation of the upholstery and cautionary tags were included, as was a sheet entitled "How to Open a Wheelchair" which gives step by step instructi
Quantity507 wheelchairs
DistributionClass II Recall - Worldwide Distribution --- including USA and Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K051302EXCEL BARIATRIC OR EXCEL SHUTTLE WHEELCHAIR
Medline Industries, Inc., Aug 15, 2005