Recall Z-2053-2008 ยท LLZ System, Image Processing, Radiological

Philips Medical Systems North America Co. Phillips

EasyVision RAD - Radiological Image Processor, software versions: SW R4 2V2L2, SW R4 2V2L3, SW R4 2V2L4, SW R4 2V2L5, SW R4 2V2L6, and SW R4 2V2L7. The EasyVision RAD consists of a computer with digital interfaces, one or two (optional) monitors with keyboard and mouse, and software. The viewing, analysis, processing and printing functions are sent to the EasyVision by means of PMSNet (Philips M

DateSep 16, 2008 (initiated Mar 10, 2008)
StatusTerminated
ReasonInaccurate measurements when exporting radiography images to Picture Archiving and Communication System (PACS ).
Root causeDevice Design
ActionOn 03/14/08 the firmed mailed certified letters to the customers of the device. The letter advises of the problem and stated that images exported from the Easy Vision RAD workstation should not be used for measurements. The letter also states that the firm will be performing upgrades of the devices. If you have questions, contact Sarah Baxter at 425-487-7665 or the Philips Call Center at 800-722-9377, #5, #2.
Quantity401 units
DistributionProducts were distributed nationwide to 159 hospitals, 3 VA's and 12 Department of Defense accounts.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K920950PHILIPS EASYVISION
Philips Medical Systems, Inc., Apr 27, 1992
K963980PHILIPS EASYVISION
Philips Medical Systems North America, Inc., Dec 23, 1996