Recall Z-1705-2008 ยท JAA System, X-Ray, Fluoroscopic, Image-Intensified
OEC Medical Systems, Inc
OEC 9800 Image Intensified fluoroscopic x-ray system (includes OEC 9800, OEC FluoroTrak 9800 Plus, OEC 9800 Plus, and OEC 9800MD Motorized C-arm System), GE Healthcare, Surgery, Salt Lake City, UT.
| Date | Sep 1, 2008 (initiated Apr 15, 2008) |
| Status | Terminated |
| Reason | Use of existing four-pedal footswitch on a different machine may cause various operational errors. |
| Root cause | Device Design |
| Action | Consignees were notified by letter on 03/21/2007 and advised that GE would be replacing the 4-pedal switches at some time in the future. Replacement began on 04/15/2008. Contact GE Healthcare at 1-800-874-7378 for assistance. |
| Quantity | 78 units |
| Distribution | Nationwide, including one VA facility in FL. No military or foreign distribution. |